Calibrator --- The Error Rate Of Annual Re-examination Is Less Than 1%, No Need To Be Re-calibrated !

May 23, 2020


  Recently, a PN-53 axial torque meter of our company's went to Shandong Institute of Metrology for a routine re-examination once a year. The machine was sold to a customer in 2017 and has been used for more than 3 years. This verification is the third verification since the machine was sold.


Shandong Institute of Metrology


Inspection Unite :  Shandong Institute of Metrology 


 axial torque meter -- Brand


The Electric Tension & Torque Calibrator PN-53 is being re-calibrated.


 axial torque meter -- Brand


The Electric Tension & Torque Calibrator PN-53 is being re-calibrated.


 axial torque meter -- Brand


The Electric Tension & Torque Calibrator PN-53 is being re-calibrated.


 axial torque meter -- Brand


We re-calibrated separately 400N.M , 800N.M , 1200N.M, 1600N.M and 2000N.M total 5 sets torque value.


 axial torque meter -- Brand


 For standard 800N.M , our machine shows 800.5N.M ;


 axial torque meter -- Brand


 For standard 1200N.M , our machine shows 1200.7N.M ;


 axial torque meter -- Brand


 For standard 1600N.M , our machine shows 1599.3N.M ;


 axial torque meter -- Brand


 For standard 2000N.M , our machine shows 2002.2N.M .


So, the inspection results shows that the error rate of our calibrator is  within the normal range !!  There is no need to be re-calibrated!